Fluke 1504 Thermometer User Manual


 
SA=5:00<EOS> sets the sample period to 5 minutes
SA=1:00:00<EOS> sets the sample period to 1 hour
7.3.1.3 Selecting the Unit of Measurement
The selected unit is used in displaying measurements on the front panel and in
reading measurements from the communications interfaces. The following
commands can be used to select the unit of measurement:
U=C<EOS> selects Celsius
U=F<EOS> selects Fahrenheit
U=K<EOS> selects Kelvin
U=O<EOS> selects ohms
7.3.1.4 Enabling the Time Stamp
Enabling the time stamp causes the time of the system clock to be transmitted
along with measurement data. The time stamp can be enabled or disabled using
the following commands:
ST=ON<EOS> enables the time stamp
ST=OFF<EOS> disables the time stamp
7.3.1.5 Setting the Clock
The system clock is set in 24-hour format using the command:
CL=hh:mm:ss<EOS>
For example:
CL=14:24:00 sets the time to 2:24 pm.
7.3.2 Probe Characterization Commands
The following commands relate to reading measurement data.
7.3.2.1 Selecting the Characterization
The following commands can be used to select the probe characterization and
coefficients:
P=T<EOS> selects the thermistor characterization
P=R<EOS> or RTD selects the Callendar-Van Dusen characterization
R0=<value><EOS> sets R
0
AL=<value><EOS> sets α for the Callendar-Van Dusen characterization
DE=<value><EOS> sets δ for the Callendar-Van Dusen characterization
BE=<value><EOS> sets β for the Callendar-Van Dusen characterization
31