Fluke 1504 Thermometer User Manual


 
29
7 Digital Communications Interface
Remote Commands
Command Description
Measurement Commands
T read measurement (includes label, unit, and time)
F[ETCH?] read measurement value (SCPI compatible)
SA[=[[[hh:]mm:]ss] read [or set] serial sample period
U=C|F|K|O select units
ST[=ON/OF] read [or set] the time stamp
CL[=hh:mm:ss] read [or set] the system clock
Probe Characterization Commands
PR[=T/R|R|] read [or select] the characterization type
R0[=<value>] read [or set] R0
AL[=<value>]
read [or set]
α
for the Callendar-Van Dusen characterization
DE[=<value>]
read [or set]
δ
for the Callendar-Van Dusen characterization
BE[=<value>]
red [or set]
β
for the Callendar-Van Dusen characterization
B
n
[=<value>] read [or set]
b
0
,b
1
,b
2
,
or
b
3
for the thermistor characterization
CO=<value> test resistance to temperature conversion
Sample Parameter Commands
FI[=<value>] read [or set] filter time constant
PS[=<value.] read [or set] the power saver period
Communication parameter commands
DU[=F/H] read [or set] serial sample duplex mode
LF[=ON/OF] read [or set] serial linefeed
Calibration Commands
*PA=<password> disable password lockout of calibration commands
*LO=[=CA|AL] read [or set] menu lockout
*C0[=<value>]
read [or set] the 0
Ω
calibration parameter
*C1[=<value>]
read [or set] the 10K
Ω
calibration parameter
*C2[=<value>]
read [or set] the 100
ΚΩ
calibration parameter
*SN[=<value>] read [or set] the instrument serial number
Miscellaneous Commands
Table 5 Command List