SYSTEMS DESCRIPTIONS
Electrical (Continued)
Drift In Microswitch (Cam 1 - All RA300 Models)
Turn Cam #1 counter-clockwise to start Drift In function
sooner (shut pump off).
Turn Cam #1 clockwise to start Drift In function later (allow
ramp to stow further before pump shuts off).
Drift Out Microswitch (Cam 2 - All RA300 Models)
Turn Cam #2 clockwise to start Drift Out function sooner
(shut pump off).
Turn Cam #2 counter-clockwise to start Drift Out function
later (allow ramp to deploy further before pump shuts off).
Figure M
Figure N
Counter Microswitch (Cam 3 - All RA300 Models)
Turn Cam #3 clockwise to start Interlock sooner.
Turn Cam #3 counterclockwise to return Interlock signal
later.
Figure L
3
2
1
3
2
1
1
2
3
Power Down Microswitch (Cam 3 - "YP" Models Only)
Turn Cam #3 clockwise to begin power down sooner.
Turn Cam #3 counterclockwise to delay power down.