Bayer HealthCare ESPRIT Blood Glucose Meter User Manual


 
CHAPTER
Technical Information 15
37
Glucometer
®
ESPRIT System
(Whole Blood Referenced)
1. Accuracy: A patient-use clinical study was
performed at diabetes care centers in three
different locations. Glucose levels were measured
with the Glucometer E
SPRIT System on over
300 fresh capillary blood specimens by 61 persons
with diabetes, and six healthcare professionals.
The results were compared with the whole blood
glucose results of the YSI 2300 Stat Plus (Yellow
Springs Instrument Co., Inc., Yellow Springs, OH),
a glucose oxidase/electrode system.
Figure 1
on the following page shows the
comparison of the lay diabetic Glucometer
E
SPRIT System results with the whole blood
glucose results. The comparison of the healthcare
professional results with the whole blood glucose
results is shown in
Figure 2
.
36
CHAPTER
15 Technical Information
REPLACEMENT PRODUCTS:
When calling or writing for supplies, be sure to include the
number with the name of the replacement part or product.
Part Number
Item
40030032 3-volt Lithium Battery (DL or
CR 2016) — 2 (Available at most stores)
99939679 Glucometer
®
ESPRIT™ User Guide
50184147 Glucometer
®
ESPRIT™ Quick Reference Guide
Part Numbers subject to change without notice.
Product
Number Product Name
3616 Glucometer
®
ESPRIT™ Test Sensor Discs — 50s
3649 Glucometer
®
ESPRIT™ Low Control
3650 Glucometer
®
ESPRIT™ Normal Control
3651 Glucometer
®
ESPRIT™ High Control
6541 Microlet™ Automatic Lancing Device
5970 Glucolanz™ Lancets (200s)
6555 Microlet™ Endcaps — Regular (10s)
6559 Microlet™ Endcaps — Super (10s)
Replacement products may be obtained from Diabetes
Care Centers, or your local pharmacy.