Bayer HealthCare Blood Glucose Meter Blood Glucose Meter User Manual


 
Need Help? Call: 0845 600 6030 (UK)/1 890 920 111 (Ireland)
Help
19
www.ascensia.co.uk
Control Solution Testing
Gently rock the control bottle
before opening to ensure it is
mixed well. Squeeze a small drop
of control solution on a clean
nonabsorbent surface (such as
a piece of wax paper). Do not
apply control solution to the
test strip directly from the
bottle.
Immediately touch the tip of
the test strip to the drop of
control solution. The solution
is pulled into the test strip
through the tip.
Hold it in the drop until the
meter beeps.
After the beep, you will see the
meter countdown five seconds
until the test is complete and your
control test result is displayed.
The meter will automatically
recognize and mark the control
result for you.
NOTE: The
does not indicate
the control test result is within
the specifi ed range.