19. SPECIFICATIONS
FX-120
i
FX-200
i
FX-300
i
FX-1200
i
FX-2000
i
FX-3000
i
Weighing capacity
122 g 220 g 320 g 1220 g 2200 g 3200 g
Maximum display
122.084 g 220.084 g 320.084 g 1220.84 g 2200.84 g 3200.84 g
Minimum weighing value (1 digit)
0.001 g 0.01 g
Repeatability (Standard deviation)
0.001 g 0.01 g
Linearity
±
0.002 g
±
0.02 g
Stabilization time (typical at FAST)
Approx. 1 second
Sensitivity drift
(10°C-30°C/50
°
F-86
°
F )
±2 ppm/°C
Operating environment
5°C to 40°C (41°F to 104°F)
85%RH or less (No condensation)
Display refresh rate
5 times/second, 10 times/second or 20 times/second
Minimum unit mass
0.001 g 0.01 g
Counting
mode
Number of samples
5, 10, 25, 50 or 100 pieces
Minimum 100% reference mass
0.100 g 1.00 g
Percent
mode
Minimum 100% display
0.01%, 0.1%, 1% (Depends on the reference mass stored.)
External calibration weight
100 g
50 g
200 g
100 g
300 g
200 g
100 g
1000 g
500 g
2000 g
1000 g
3000 g
2000 g
1000 g
Weighing pan diameter
130 mm 150 mm
Net weight
Approx. 2.5 kg
External dimensions
193 (W) x 262.5 (D) x 84.5 (H) mm
AC adapter
Confirm that the adapter type is correct for the local voltage and power
receptacle type
Power consumption
Approx. 11VA (supplied to the AC adapter)
Interface
RS-232C