A&D FX-120i/FX-200i/FX-300i/FX1200i/FX2000i/FX3000i Scale User Manual


 
51
15. COMMANDS
15-1 Command List
Note
A command has a terminator added, that is specified using the "Terminator (Crlf)"
parameter of “Serial interface ( 5if )” in the function table, and is sent to the balance.
Commands to query weight data
C Cancels the S or SIR command.
Q Requests the weight data immediately.
S Requests the weight data when stabilized.
SI Requests the weight data immediately.
SIR Requests the weight data continuously.
E
S
C
P Requests the weight data when stabilized.
Note: The “Q” and “SI” commands, the “S” and “
E
S
C
P” commands behave the same.
Commands to control the balance
CAL
Same as the CAL key.
OFF Turns the display off.
ON Turns the display on.
P
Same as the ON:OFF key
PRT
Same as the PRINT key
R
Same as the RE-ZERO key
SMP
Same as the SAMPLE key.
T Tare key
Z
Same as the RE-ZERO key
E
S
C
T
Same as the RE-ZERO key
U
Same as the MODE key
?ID Requests the ID number.
?SN Requests the serial number.
?TN Requests the model name.
?PT Requests the tare weight.
PT: ∗∗∗∗∗. g
Changes the tare weight.
The unit added is the current weighing unit in A&D standard format.
Note: The “R”, “Z” and “
E
S
C
T” commands behave the same.
E
S
C
:
ASCII code 1Bh