A&D FX-120i/FX-200i/FX-300i/FX1200i/FX2000i/FX3000i Scale User Manual


 
24
7-2 Calibration Using an External Weight
This function calibrates the balance using an external weight.
Operation
1 Connect the AC adapter and warm up the balance
for at least 30 minutes with nothing on the pan.
2 Press and hold the CAL key until Calout is
displayed, and then release the key.
3 The balance displays Cal 0 .
If you want to change the calibration weight (a
list of usable weights is shown on page 23),
press the SAMPLE key and proceed to step 4.
If you use the calibration weight value stored in
the balance, proceed to step 5.
4 Specify the calibration weight value as follows:
SAMPLE key To switch the display condition
to: “All of the segments
blinking” (calibration weight
selection mode) or “The last
two digits blinking” (value
adjustment mode).
RE-ZERO key (To increase the value)
MODE key (To decrease the value)
To select the calibration
weight or adjust the value.
PRINT key To store the new weight
value. Even if the AC
adapter is removed, the data
is maintained in non-volatile
memory.
CAL key To cancel the operation and
return to Cal 0 .