18
10. ERROR MESSAGES
Overload error
Warning to indicate that an object beyond the weighing
capacity has been placed on the weighing pan. Remove
the object from the weighing pan.
Power-on error
Warning to indicate that an object beyond the power-on
zero range (50% of the weighing capacity or 10%) is
placed on the weighing platform. Remove the object from
the weighing pan.
Pressing the PRINT/ENT key, the indicator will show
the weight value without zeroing the indicator.
Low battery
Warning to show that the OP-02 battery is discharged.
Immediately connect the power cord to the AC power
source to charge the battery. The indicator can be used
while the battery is charging (except when the OP-04 is
installed).
CAL error
Warning to indicate that the calibration has been
canceled because the calibration weight is too light.
Check if the calibration weight is correct.
Other errors
Showing errors inside the indicator. Check the
connection between the indicator and platform, and turn
the power off and on with nothing on the weighing
platform. (# signifies an error number.)
When the indicator does not recover from errors or shows other errors, contact your
local A&D dealer.
----e---
--------
---lb1--
---Cal-e
---err-#