A&D 1WMPD4002313 Scale User Manual


 
16
The frame ground and the shield are connected internally and there is no difference
between them in use.
Data format
Example of data
Out of range “kg” (+)
O L
,
+999999 9 9 _ k g C
R
L
F
Out of range “kg” (+)
O L
,
+9999
.
9 9 9 _ k g C
R
L
F
The position of decimal point is different according to the model and/or “f02”.
Example for the weighing unit “lb-oz” (added)
Weighing data “lb-oz” (+)
S T
,
+001L01
.
6 _ o z C
R
L
F
Data output mode (f06)
The commands can be used in all data output modes (f06-1 is command mode only).
The new modes f06-5, f06-6 and f06-7 are added to the Function “f06”.
Auto-print Mode + data and OK (f06-6)
Data is sent if the weight display is stable at +5d (d = weighing display division) and
above, and the comparator result OK in addition. The next transmission cannot
occur until after the weight display falls below +5d.
Auto-print Mode +/- data and OK (f06-7)
Data is sent if the weight display is stable at ±5d (d = weighing display division) and
above/below, and the comparator result OK in addition. The next transmission
cannot occur until after the weight display falls between 5d and +5d.
Multi-connection with Print key mode (RS-422/485) (f06-5)
This is one of the applications when a PC and more than one FS-D (RS-422/485
installed) are connected. Refer to “Multi-connection with Print key mode” in detail.
(In “2-3. Command Mode”)
Examples of command and reply (“
_
” shows “Space” (20H).)
There are corrections to the instruction manual 1WMPD4001368.
Set a HI limit value or upper limit weight. (No reply for f20-1.)
When f07-0 or f07-1 is set, the
command should have “+” and 6
digit number without decimal point.
When f07-0 or f07-1 is set, the command
should have “+/-” (“+” only for f07-1) and 6
digit number without decimal point.
Set a LO limit value or lower limit weight. (No reply for f20-1.)
When f07-0 or f07-1 is set, the
command should have “+” and 6
digit number without decimal point.
When f07-0 or f07-1 is set, the command
should have “+/-” (“+” only for f07-1) and 6
digit number without decimal point.
When f07-2 is set, the command should have “+” and 5 digit number without
decimal point.