The frame ground and the shield are connected internally and there is no difference
between them in use.
Data format
Example of data
Out of range “kg” (+)
O L
,
+999999 9 9 _ k g C
R
L
F
Out of range “kg” (+)
O L
,
+9999
.
9 9 9 _ k g C
R
L
F
The position of decimal point is different according to the model and/or “f02”.
Example for the weighing unit “lb-oz” (added)
Weighing data “lb-oz” (+)
S T
,
+001L01
.
6 _ o z C
R
L
F
Data output mode (f06)
The commands can be used in all data output modes (f06-1 is command mode only).
The new modes f06-5, f06-6 and f06-7 are added to the Function “f06”.
Auto-print Mode + data and OK (f06-6)
Data is sent if the weight display is stable at +5d (d = weighing display division) and
above, and the comparator result OK in addition. The next transmission cannot
occur until after the weight display falls below +5d.
Auto-print Mode +/- data and OK (f06-7)
Data is sent if the weight display is stable at ±5d (d = weighing display division) and
above/below, and the comparator result OK in addition. The next transmission
cannot occur until after the weight display falls between –5d and +5d.
Multi-connection with Print key mode (RS-422/485) (f06-5)
This is one of the applications when a PC and more than one FS-D (RS-422/485
installed) are connected. Refer to “Multi-connection with Print key mode” in detail.
(In “2-3. Command Mode”)
Examples of command and reply (“
_
” shows “Space” (20H).)
There are corrections to the instruction manual 1WMPD4001368.
Set a HI limit value or upper limit weight. (No reply for f20-1.)
When f07-0 or f07-1 is set, the
command should have “+” and 6
digit number without decimal point.
When f07-0 or f07-1 is set, the command
should have “+/-” (“+” only for f07-1) and 6
digit number without decimal point.
Set a LO limit value or lower limit weight. (No reply for f20-1.)
When f07-0 or f07-1 is set, the
command should have “+” and 6
digit number without decimal point.
When f07-0 or f07-1 is set, the command
should have “+/-” (“+” only for f07-1) and 6
digit number without decimal point.
When f07-2 is set, the command should have “+” and 5 digit number without
decimal point.