A&D 1WMPD4002313 Scale User Manual


 
11
6-2. Function list
Item
Function number
Description
C01- 0 CAL switch enabled, 0 key enabled
C01- 1 CAL switch disabled, 0 key disabled
Calibration
C01- 2 CAL switch enabled, 0 key disabled
Do not set
C01-1
C02- 0 6 kg (15 lb)
C02- 1 15 kg (35 lb)
C02- 2 30 kg (70 lb)
C02- 3 60 kg (150 lb)
C02- 4 150 kg (350 lb)
C02- 5 300 kg (700 lb)
C02- 6 600 kg (1500 lb)
Capacity
C02- 7 1500 kg (3500 lb)
C03- 0 kg, g
Weighing display
unit
C03- 1 kg, g, lb, oz, lb-oz
See “11-1.
Capacity ...”.
C04- 0 ±10% of the capacity OIML R76 / NTEP
Power-on-zero
range
C04- 1 ±50% of the capacity
Internal use C05- 0 Do not change.
C06- 0 . (point)
Decimal
separator
C06- 1 , (comma)
C07- 0 ZERO does not clear tare weight
Tare clear by
ZERO operation
C07- 1 ZERO clears tare weight OIML R76
C08- 0
Tare weight is rounded to the weight
display count
NTEP
Tare operation
(Net weight
calculation)
C08- 1 Internal count is used for TARE weight. OIML R76
C09- 0 Function disabled
Preset tare
function
C09- 1 Function enabled
C10- 0 lb-oz disabled
lb-oz unit
C10- 1 lb-oz enabled
C11- 0 ±2% of the capacity
ZERO operation
range
C11- 1 Up to the capacity
C12- 0 f02” setting specifies the resolution.
C12- 1 Fixed to “f02-0” (1/3,000~1/3,500)
C12- 2 Fixed to “f02-1” (1/6,000~1/7,500)
Display resolution
C12- 3 Fixed to “f02-2” (1/12,000~1/15,000)
C13- 0 Function disabled
Auto-tare function
C13- 1 Function enabled
C13-0 deletes
f21” ~ “f23”.
Factory setting