Commands by Subsystem 3
Agilent InfiniiVision 5000 Series Oscilloscopes Programmer's Reference 353
:TRIGger:GLITch:RANGe
(see page 530)
Command Syntax
:TRIGger:GLITch:RANGe <greater than time>[suffix],
<less than time>[suffix]
<greater than time> ::= start time (10 ns - 9.99 seconds) in NR3 format
<less than time> ::= stop time (15 ns - 10 seconds) in NR3 format
[suffix] ::= {s | ms | us | ns | ps}
The :TRIGger:GLITch:RANGe command sets the pulse width duration for
the selected :TRIGger:GLITch:SOURce. If you set the stop time before the
start time, the order of the parameters is automatically reversed.
Query Syntax
:TRIGger:GLITch:RANGe?
The :TRIGger:GLITch:RANGe? query returns the pulse width duration time
for :TRIGger:GLITch:SOURce.
Return Format
<start time>,<stop time><NL>
See Also • "Introduction to :TRIGger Commands" on page 313
• ":TRIGger:GLITch:SOURce" on page 354
• ":TRIGger:GLITch:QUALifier" on page 352
• ":TRIGger:MODE" on page 319