Tektronix 544A Blood Glucose Meter User Manual


 
Waveform Integration
Reference
6. Read the integrated voltage over time between the cursors in volt-
seconds from the : readout. Read the integrated voltage over time
between the selected cursor and the reference indicator of the math
waveform from the @: readout. See Figure 3-83.
Source
Waveform
Integral Math
Waveform
Figure 3-83: H Bars Cursors Measure an Integral Math Waveform
7. Press Function (main) V Bars (side). Use the general purpose knob
to align one of the two vertical cursors to a point of interest along the
horizontal axis of the waveform.
8. Press SELECT to select the alternate cursor.
9. Align the alternate cursor to another point of interest on the math wave-
form.
10. Read the time difference between the cursors from the : readout. Read
the time difference between the selected cursor and the trigger point for
the
source
waveform from the @: readout.
11. Press Function (main) Paired (side).
12. Use the technique just outlined to place the long vertical bar of each
paired cursor to the points along the horizontal axis you are interested in.
13. Read the following values from the cursor readouts:
Read the integrated voltage over time between the
X
s of both paired
cursors in volt-seconds from the : readout.