WARNINGS
Read the entire manual before using the device.
Use the device only as directed by your physician or healthcare provider.
Use the device only for the intended use as described in this manual. Advice contained in this
manual should not supersede instructions given by the prescribing physician.
If you notice any unexplained changes in the performance of the device, if it is making unusual
or harsh sounds, if the device or the power supply are dropped or mishandled, if water is spilled
into the enclosure, or if the enclosure is broken, discontinue use and contact your ResMed
Service Center.
Beware of electrocution. Do not immerse the device, humidifier, power supply or power cord in
water. In the event of a spill, disconnect the device from the power supply and let the parts dry.
Always unplug the device before cleaning and make sure that all parts are dry before plugging in
the device.
Explosion hazard—do not use in the vicinity of flammable anesthetics.
Make sure the power cord and plug are in good condition and the equipment is not damaged.
Keep the power cord away from hot surfaces.
The device should only be used with masks (and connectors
1
) recommended by ResMed, or by
a physician or respiratory therapist. A mask should not be used unless the device is turned on.
Once the mask is fitted, ensure that the device is blowing air. The vent hole or holes associated
with the mask should never be blocked.
Explanation: The device is intended to be used with special masks (or connectors) which have
vent holes to allow continuous flow of air out of the mask. When the device is turned on and
functioning properly, new air from the device flushes the exhaled air out through the mask vent
holes. However, when the device is not operating, insufficient fresh air will be provided through
the mask, and the exhaled air may be rebreathed. Rebreathing of exhaled air for longer than
several minutes can, in some circumstances, lead to suffocation. This applies to most models of
CPAP or bilevel devices.
Oxygen supports combustion. Oxygen must not be used while smoking or in the presence of an
open flame.
Always ensure that the device is turned on and airflow generated before the oxygen supply is
turned on. Always turn the oxygen supply off before the device is turned off, so that unused
oxygen does not accumulate within the device enclosure and create a risk of fire.