Air tubing technical specifications
Air tubing Material Length Inner diameter
ClimateLine heated air tubing Flexible plastic and electrical
components
2 m 15 mm
ClimateLine
MAX
heated air
tubing
Flexible plastic and electrical
components
1.9 m 19 mm
SlimLine air tubing Flexible plastic 1.8 m 15 mm
Standard air tubing Flexible plastic 2 m 19 mm
3 m air tubing Flexible plastic 3 m 19 mm
Heated air tubing temperature cut-out:
41°C
Notes:
The manufacturer reserves the right to change these specifications without notice.
The temperature and relative humidity settings displayed for Climate Control are not measured
values.
Check with your clinician/service provider before using the SlimLine air tubing with devices other
than the S9 or H5i.
The electrical connector end of the heated air tubing is only compatible with the H5i air outlet
and should not be fitted to the device or mask.
When using the SlimLine or ClimateLine above 20 cm H
2
O, the device optimum performance
may not be reached if used with an antibacterial filter. The device performance must be checked
prior to prescribing the SlimLine for use with an antibacterial filter.
The ClimateLine or ClimateLine
MAX
is designed only for use with the H5i.