Quantum 610 Mobility Aid User Manual


 
Quantum 610 Series www.pridemobility.com 37
FRONT RIM
HALF
WHEEL
HUB
VII. CARE AND MAINTENANCE
Figure 27. Quantum 610 Drive Wheel Disassembled
WASHERS
TIRE
REAR RIM HALF
NUTS
Figure 26. Quantum 610 Drive Wheel
TUBE
DRIVE WHEEL
WASHER
DRIVE
WHEEL
NUT
AXLE
SLOT
AXLE KEY
Follow these easy steps for a quick and safe repair for both pneumatic and solid tires:
1. Turn off the power to the controller.
2. Set the power chair up on blocks.
3. If you are changing a pneumatic tire, completely deflate it before removing the wheel.
4. Remove the drive wheel nut and washer from the axle. See figure 26.
5. Pull the wheel off the axle.
6. Remove the nuts and washers from the wheel hub and separate the rim halves. See figure 27.
7. Remove the old tube from the pneumatic tire and replace it with a new tube or replace the entire assembly if
it is a solid tire.
8. Reassemble the rim halves and reinstall the nuts and washers to the wheel hub.
9. Slide the wheel back onto the axle. Make sure that the key is in the axle slot.
WARNING! Failure to ensure that the axle key is properly installed into the axle slot when
mounting the wheel can result in electronic brake failure, personal injury, and/or product damage.
10. Reinstall the drive wheel nut and washer to the axle and tighten.
WARNING! Make sure both the nut and washer are reinstalled and secured properly.
11. Inflate pneumatic tires to the psi/bar/kPa air pressure rating indicated on each tire.
12. Remove the power chair from the blocks.
Battery Replacement
A battery wiring diagram is printed on a decal located on the front battery tray. See VI. “Batteries and Charging”
for correct battery specifications.
MANDATORY! Battery posts, terminals, and related accessories contain lead and lead
compounds. Wear goggles and gloves when handling batteries and wash hands after handling.
WARNING! The batteries in your power chair should only be serviced or replaced by a
Quantum Rehab Provider or a qualified technician.
WARNING! Do not replace batteries when seat is occupied.
WARNING! Power chair batteries are heavy. See specifications table. If you are unable
to lift that much weight, be sure to get help. Use proper lifting techniques and avoid
lifting beyond your capacity.
38 www.pridemobility.com Quantum 610 Series
VII. CARE AND MAINTENANCE
WARNING! Do not mix old and new batteries. Always replace both batteries at the same time.
PROHIBITED! Keep tools and other metal objects away from the battery terminals. Contact with
tools can cause electrical shock.
To replace the batteries:
1. Turn off the power to the controller.
2. Make sure that the power chair is in drive mode. See III. “Your Power Chair.”
3. Loosen the rear shroud fasteners.
4. Remove the rear shroud. See figure 28.
5. Disconnect the battery harness from the rear battery.
6. Remove the retaining pin and lift off the front cover using the foot platform bracket. See figure 20.
7. Slide the battery tray forward.
8. Disconnect the battery harness from the front battery.
9. Unfasten the hook and loop straps from both batteries.
10. Remove the batteries.
11. Install new batteries.
12. Fasten the hook and loop straps around both batteries.
13. Connect the battery harness to the front battery according to the battery wiring diagram. See figure 28.
WARNING! Make sure you tighten the fasteners so that the connections are secure.
14. Slide the battery tray back into the power base.
15. Connect the battery harness to the rear battery.
16. Reinstall the front cover.
17. Reinstall the rear shroud and tighten the fasteners.
18. Charge the batteries. See VI. “Batteries and Charging.”
When to See Your Quantum Rehab Provider for Service
The following symptoms could indicate a serious problem with your power chair. If necessary, contact your
Quantum Rehab Provider. When calling, have the model number, serial number, nature of the problem, and the
error code if available.
Motor noise
Frayed harnesses
Cracked or broken connectors
Uneven wear on any of the tires
Jerky motion
Pulling to one side
Bent or broken wheel assemblies
Does not power up
Powers up, but does not move
Loose seat or seating components
Corrective Maintenance
If the battery condition meter does not light up when you turn on the power:
Check the harness connections. Make sure they are tight.
Check the circuit breaker. Reset it if necessary.
Check the battery connections.
If the above conditions prove normal, you can load test the batteries with a battery load tester. These testers are
available at automotive parts stores. Disconnect both batteries before load testing and follow the directions that
come with the load tester. If either one of the batteries fails the load test, replace both of them. If your power chair
still does not power up, contact your authorized Quantum Rehab Provider.