Pride Mobility 6000Z 3MP Mobility Aid User Manual


 
Quantum 6000Z Series www.quantumrehab.com 37
VII. CARE AND MAINTENANCE
! The body shroud has been sprayed with a clear sealant coating. You can apply a light coat of car wax to help it retain
its high-gloss appearance.
! Check all electrical connections. Make sure they are tight and are not corroded. Batteries must sit flat within the battery
boxes and battery well frame with the battery terminals facing inward toward each other. Refer to the battery wiring
label for the correct wiring layout.
WARNING! Even though the power chair has passed the necessary testing requirements
for ingress of liquids, you should keep electrical connections away from sources of
dampness, including direct exposure to water or bodily fluids and incontinence. Check
electrical components frequently for signs of corrosion and replace as necessary.
! All wheel bearings are prelubricated and sealed. They require no subsequent lubrication.
Daily Checks
! With the controller turned off, check the joystick. Make sure it is not bent or damaged and that it returns to the neutral
position when you release it. Check the rubber boot around the base of the joystick for damage. Visually inspect the boot.
Do not handle or try to repair it. See your Quantum Rehab Specialist if there is a problem.
! Visually inspect the controller communications cable. Make sure that it is not frayed, cut, or has any wires exposed.
See your Quantum Rehab Specialist if there is a problem.
! Check for flat spots on solid tires. Flat spots could adversely affect stability.
Weekly Checks
! Disconnect and inspect the controller cable and the charger harness from the power base. Look for corrosion. Contact
your Quantum Rehab Specialist if necessary.
! Ensure that all parts of the controller system are securely fastened to your power chair. Do not overtighten any screws.
! Check for proper tire inflation. There should be 35 psi (2.4 bar) in each tire. If a tire does not hold air, see your
Quantum Rehab Specialist for replacement of the tube.
! Check the brakes. This test should be carried out on a level surface with at least 3 feet (1 meter) of clearance around your
power chair.
To check the brakes:
1. Turn on the controller and turn down the speed level of your power chair.
2. After one second, check the battery condition meter. Make sure that it remains on.
3. Slowly push the joystick forward until you hear the electric brakes click. Immediately release the joystick. You must be
able to hear each electrical brake operating within a few seconds of joystick movement. Repeat this test three times,
pushing the joystick backward, then left, and then right.
Monthly Checks
! Check for drive tire wear. See your Quantum Rehab Specialist for repair.
! Check the caster wheels for wear. Replace them as necessary.
! Check the caster forks for damage or fluttering which indicates that they may need to be adjusted or have the bearings
replaced. See your Quantum Rehab Specialist for repair.
! Keep your power chair clean and free of foreign material, such as mud, dirt, hair, food, drink, etc.
Yearly Checks
Take your power chair to your Quantum Rehab Specialist for yearly maintenance, especially if you use your power chair
on a daily basis. This helps ensure that your power chair is functioning properly and helps prevent future complications.