Pride Mobility 400 Mobility Aid User Manual


 
16 www.pridemobility.com Commander 400
VI. CARE AND MAINTENANCE
The following routine maintenance should be performed at the indicated intervals. Failure to perform the
required maintenance procedures at the required intervals may void Pride’s product warranty.
Perform prior to daily use:
Inspect lift strap wear and replace if necessary.
Inspect all harnessing for damage and replace if necessary.
Perform every 6 months:
Inspect all mounting hardware for defects and a secure fit. Any hardware found to be loose should be
tightened or replaced if necessary.
Perform every 6 – 12 months:
Visually inspect chains for damage and replace if necessary.
Lubricate chains with “General Purpose Petroleum-based Grease”.
Inspect plastic shrouds for cracks or damage and replace if necessary.
Inspect the power base plastic shrouds for cracks or damage and replace if necessary.
Lubricate the power base gears with “General Purpose Petroleum-based Grease”.
Storage and Operation Temperatures
Transportation or storage: -40°F/-40°C to 149°F/65°C
Operation: -13°F/-25°C to 122°F/50°C