LIFT WILL NOT OPERATE
Check the circuit breaker reset button. Reset if
necessary. See figure 6.
If the circuit breaker reset button is not tripped,
check the hardwire fuse to see if it is blown.
Replace if necessary. See figure 9.
Ensure all harness connections are secure.
Ensure the power wire leading to the positive
battery terminal is tight and not corroded.
Check the fuse located on the red power wire
near the battery. If blown, replace with a new
15-amp fuse. See figure 9.
WARNING! The replacement fuse
must exactly match the rating of
the old fuse. Failure to use properly
rated fuses may cause damage to
the electrical system.
Figure 9. Hardwire Fuse
V. TROUBLESHOOTING
LIFT STOPS OPERATING
If your lift system is heavily strained because of excessive loads, the main circuit breaker (see figure 6)
may trip to protect the motor and electronics from damage. When the breaker trips, the entire electrical
system shuts down.
WARNING! Never attempt to lift more than 400 lbs. (181 kg) with the lift system.
To reset the main circuit breaker:
1. Allow a minute or so for the electronics to “cool.”
2. Push in the reset button to reset the breaker. See figure 6.
If the breaker trips frequently, contact your authorized Pride Provider.
LIFTING STRAP TRAVELS IN OPPOSITE DIRECTION OF BUTTON BEING
PUSHED ON HAND CONTROL
Before each use, check that the lifting strap is traveling in the proper direction. If the belt travels in the
wrong direction when you press either the up or down button, you must restore proper lifting strap travel.
To restore proper lifting strap travel:
1. Pull slightly on the strap and operate the lift until the strap is extended completely past the “stop”
warning label.
2. Allow the strap to rewind itself. Do not allow any twists in the belt.
Observe the travel of the belt, which should correspond with the button being pushed on the hand control.
If you are still experiencing problems, contact your authorized Pride Provider.
WORKING FUSE
BLOWN FUSE
15 AMP