Philips M5066A Blood Pressure Monitor User Manual


 
G-4
Philips Medical Systems
HEARTSTART M5066A
Troubleshooting while the HeartStart is not in use
(green Ready light is not on)
behavior possible cause recommended action
chirps or
i-button flashes
The battery power is low or the
SMART Pads cartridge needs to be
replaced.
The defibrillator may have been
turned off without a pads cartridge
installed, or the installed pads
cartridge may not have its hard
cover in place.
The training pads cartridge has been
left in the defibrillator.
The defibrillator has been stored
outside the recommended
temperature range.
The defibrillator has detected an
error during a self-test or cannot
perform a self-test, or the Shock
button is damaged.
Press the blue i-button. Replace the
battery or pads cartridge if
instructed.
Make sure the pads cartridge is
properly installed with the hard
cover in place. (See Chapter 5,
“Maintaining the HeartStart,” for
directions on installing the pads
cartridge.)
Remove the training pads cartridge
and replace it with an Adult or
Infant/Child Pads Cartridge.
Remove the battery for five
seconds then reinstall it to start
the battery insertion self-test. If it
fails, insert a new battery to repeat
the test. If it fails again, do not use
the defibrillator. If it passes, store
the defibrillator within the
recommended temperature range.
Contact Philips for service if
needed.
no chirping and/or
i-button does not flash
The battery is missing or completely
depleted.
The defibrillator may have been
physically damaged.
Remove the battery for five seconds
then reinstall it to start the battery
insertion self-test. If it fails, insert a
new battery and repeat the test. If it
fails again, do not use the defibrillator.
Contact Philips for service.