Philips M5066A Blood Pressure Monitor User Manual


 
Philips Medical Systems
5-1
5
5 Maintaining the HeartStart
Routine Maintenance
The HeartStart is very simple to maintain. The defibrillator performs a self-test
every day. In addition, a battery insertion self-test is run whenever a battery is
installed in the device. The defibrillator’s extensive automatic self-test features
eliminate the need for any manual calibration. The HeartStart has no user-
serviceable parts.
WARNING: Electrical shock hazard. Do not open the HeartStart, remove its
covers, or attempt repair. There are no user-serviceable components in the
HeartStart. If repair is required, return the HeartStart to Philips for service.
Reminders:
Do not leave the HeartStart without a pads cartridge installed; the
defibrillator will start chirping and the i-button will start flashing. For
directions on changing the pads cartridge, see Chapter 2, “Setting up the
HeartStart.”
The HeartStart runs daily self-tests. As long as the green Ready light is
blinking, it is not necessary to test the defibrillator by initiating a battery
insertion self-test. This uses battery power and risks draining the battery
prematurely.
Periodic checks
Other than the checks recommended after each use of the HeartStart,
maintenance is limited to periodically checking the following:
Check the green Ready light. If the green Ready light is not blinking, see
Troubleshooting Tips, below.
Replace any used, damaged or expired supplies and accessories
Check the outside of the defibrillator. If you see cracks or other signs of
damage, contact Philips for technical support.