Power Down Sequence
1.
Return the viewing screen back to its original position
and spread the beam.
2.
If high magnification was used, lower the magnification
to about 7100x by moving the magnification knob
counter-clockwise.
3.
Go to MODE > CONFIGURATION then turn the filament
knob counter-clockwise until the filament level goes
back to zero. (Note: There is no wait time when
decreasing the filament saturation.
4.
Turn off the HIGH TENSION switch and turn off the
control panel. If another user will use the microscope
soon after, just press VACUUM from the main menu to
see the vacuum status then leave it at that.
5.
Cover the viewing windows with their lids.
6.
Logoff from the computer to lock the control panel.
7.
Sign the microscope user logsheet and indicate if the
microscope has been in good working condition.