Omega Engineering CYD201 and CYD208 Thermometer User Manual


 
OMEGA Model CYD201/CYD208 User’s Manual
5-6 Service
5.7.2 A/D Converter Calibration
NOTE:
Current source calibration
must
be performed
before
A/D converter
calibration.
1. Allow 30 minute warm-up to achieve rated specifications.
2. Perform current source calibration.
3. Configure 125 k resistor as shown in Figure 5-4. Set Model 208s to
channel 1.
4. Wait 10 seconds for resistor voltage to settle.
5. Press
CAL ENABLE
, then within 2 seconds press
UNITS
. The display
window shows "-CAL-" for approximately 15 seconds.
6. Repeat procedure with 25 k resistor.
7. For Model CYD208, repeat procedure for other 7 channels. See
Figure 2-2 for pin assignments.