Omega Engineering CYD201 and CYD208 Thermometer User Manual


 
OMEGA Model CYD201/CYD208 User’s Manual
Remote Operation 4-5
F0 Sets Sensor Units for Temperature Display.
Input:
F0x
Returned:
Nothing
Remarks:
Sets sensor units for the temperature display, where
x
=
C
(Celsius),
F
(Fahrenheit),
K
(kelvin), or
V
(volts).
H Select High Alarm Setpoint Value.
Input:
Hxxx.x
Returned:
Nothing
Remarks:
Sets high alarm setpoint, where
xxx.x
= temperature setpoint
in units specified by F0 command. If the instrument is set for
volts, the alarm defaults to kelvin.
Example: H300[term]
sets a high alarm setpoint of 300 degrees.
L Select Low Alarm Setpoint Value.
Input:
Lxxx.x
Returned:
Nothing
Remarks:
Sets low alarm setpoint, where
xxx.x
= temperature setpoint
in units specified by F0 command. If the instrument is set for
volts, the alarm defaults to kelvin.
Example: L31.2[term]
sets a low alarm setpoint of 31.2 degrees.
R Alarm Reset.
Input:
R
Returned:
Nothing
Remarks:
Resets the alarm.
Y Channel Dwell Time
(Model CYD208 Only)
.
Input:
Yab
Returned:
Nothing
Remarks:
Sets dwell time for a given channel, where
a
= channel 1 - 8,
and
b
= the dwell time parameter as follows:
0
= zero seconds,
1
= 5 seconds,
2
= 10 seconds,
3
= 30 seconds,
4
= 60 seconds
Setting a dwell time of 0 skips the specified channel in the
sequence.
Example: Y23[term]
sets the dwell time for channel 2 to 30 seconds.