OMEGA Model CYD201/CYD208 User’s Manual
Service 5-3
5.3 ERROR CODE TROUBLESHOOTING
On power up, the CYD201/CYD208 checks internal memory. If a problem
exists, an error message displays on the front panel of the instrument.
Er01
indicates a hardware problem in the instrument memory. This error is
not user-correctable.
Er02
indicates a soft error in the instrument memory. To correct this error,
use the following procedure.
1. Power up the unit and allow it to display Er02.
2. Close DIP Switch 1 (top of the switch pressed in). Leave the switch
closed for at least 5 seconds, then open DIP switch 1 (bottom of the
switch pressed in).
3. Verify the Model CYD201/CYD208 display goes through a normal power
up sequence and then displays 499.9 K.
4. The input(s) of the Model CYD201/CYD208 must now be recalibrated
per in procedure in Paragraph 5.7 before the unit can be used.
OL
indicates a voltage input overload. This can be caused by an open
sensor or diode sensor wired backwards.
Before calling the factory about a persistent problem, try the procedure
below:
WARNING: This procedure erases calibration constants stored in Non-
Volatile RAM. If this procedure is used, recalibrate the instrument.
1. With power turned Off, press and hold
CAL ENABLE
on the back panel.
While holding
CAL ENABLE
, turn instrument power On.
2. If the Model CYD201/CYD208 displays
Er02
, follow the Calibration
procedure in Paragraph 5.7. If
Er01
still displays or if the Model
CYD201/CYD208 does not respond, contact OMEGA Service.
5.4 GENERAL MAINTENANCE
Clean the CYD201/CYD208 periodically to remove dust, grease and other
contaminants. Clean the front and back panels and case with a soft cloth
dampened with a mild detergent and water solution.
NOTE:
Do not use aromatic hydrocarbons or chlorinated solvents to clean
the Model CYD201/CYD208. They may react with the silk screen printing on
the back panel.