ERROR MESSAGES
The apparatus registers any problems and shows the following error messages
on the display:
30
Error message Meaning Cause Solution
ER-1 Error during 1. The armband is not correctly attached. 1. Attach the armband correctly.
pumping 2. The armband is broken. 2. Contact the retailer.
3. The pump or valves are defective. 3. Contact the retailer.
ER-2 Disturbances or 1. The user moves or speaks 1. The user must sit still and avoid talking during the reading.
noise during readings during the reading.
2. The user is too nervous. 2. Have a blood pressure
reading taken by a doctor.
ER-3 Error while emptying 1. The valves are defective. 1. Contact the retailer.
the armband 2. The armband is not correctly attached. 2. Attach the armband correctly.
ER-4 Abnormal reading The user may have a heart problem. Consult a doctor.
HI The pressure in the The air tube is blocked, the pump Try to take another reading,
armband is too high. mechanism is defective or the pressure or contact the retailer.
monitor is defective.
Battery symbol Flat batteries The batteries are almost dead. Replace the batteries.