Important Control Solution Test Information
• Use only OneTouch
®
Ultra
®
Control Solution.
• Check the expiration date on the control solution vial.
Do not use if expired.
• Control solution, meter, and test strips should come to room
temperature (20-25°C/68-77°F) before testing.
• Shake the vial, discard the first drop of OneTouch
®
Ultra
®
Control
Solution, and wipe off the dispenser tip to ensure a good sample
and an accurate result.
• Use only for three months after first opening. Record the discard
date (date opened plus three months) on the control solution
vial. Discard after three months.
• Store the control solution tightly closed at room temperature
below 30°C (86°F ). Do not refrigerate.
CAUTION: The control solution range printed on the test strip vial
is for OneTouch
®
Ultra
®
Control Solution only. It is used to test
meter and test strip performance. It is not a recommended range
for your blood glucose level.
DCO_05265303A_OTFT_OB_CAN_EN 2/2/04 2:39 PM Page 13