ACCESSORIES
Part No. 1145810 47 Electric Portable Patient Lift
2. WiththeReliantScaleon,removethefourscrewsonthebackoftheenclosureand
removethefrontcovertoexposethePCboard.
3. PresstheCALbuttonlocatedonthePCboard.TheCALswitchisnotlabeledbutis
theonlybuttononthePCboardlocatedinthelowerrightcorner.OncetheCAL
buttonispushedʺCAL1ʺwillbeseeninthedisplaywindow.
NOTE:CAL1selectsthecalibrationmodeusing50poundsofcalibratedweight.CAL2selectsthe
calibrationmodeforusewith200poundsofcalibratedweight.PressingtheUNITSkeytoggles
betweenCAL1andCAL2.
4. Whenthedesiredcalibrationmodeisdisplayed,presstheLOCK/UNLOCKkey.
NOTE:ThedisplaywindowwillnowshowUnLd.
5. EnsurethatthereisnoloadonthescaleandpresstheZEROkey.
6. Thedisplaywindow willshowadashedline(‐‐‐‐‐‐)scrollingacrossandthentheword
LOADwillappearinthedisplaywindow.
7. Performoneofthefollowing:
A. FORCAL1OPTION‐support50poundsofcalibratedweightfromthescaleand
presstheZEROkey.
B. FORCAL2OPTION‐support200poundsofcalibratedweightfromthescaleand
presstheZEROkey.
8. Thedisplaywindow willshowadashedline(‐‐‐‐‐‐)scrollingacrossandthentheword
DONEwillappearinthedisplaywindow.
9. RemovetheweightfromthescaleandpresstheLOCK/UNLOCKkey.
NOTE:Theunitwillnowbeinatemporarytestmodeandwillnotlocktheweightdisplay.This
willallowweighttobeloadedandunloadedtocheckthecalibration.
10. TurntheunitoffbypressingtheON/OFFkey.
11. TurntheunitonbypressingtheON/OFFkey.
NOTE:Scalewillnowbeinnormaloperation.
Troubleshooting
SYMPTON PROBABLE CAUSE SOLUTIONS
Unit does NOT work properly. Battery failure. Check battery. Replace if necessary.
Battery has been replaced and unit
still does NOT work properly.
--
Contact Invacare for service at 1-800-333-
6900