Invacare R110 Mobility Aid User Manual


 
MAINTENANCE
Part No. 1145810 39 Electric Portable Patient Lift
FIGURE 7.4 Replacing the Swivel Bar
Maintaining the Base Adjustment
NOTE:Forthisprocedure,refertoFIGURE 7.5.
Thebaseadjustmentshouldnotrequireanyattentionotherthan:
1. Checkthatthelegsaresquarewhenintheclosedposition.
2. Placeasquareontheinsideofthelegsandbasetodeterminethe90°alignment.
3. Adjustthelinkagerods until90°alignmentisachieved.
FIGURE 7.5 Maintaining the Base Adjustment
600 lbs Swivel Bar
450 lbs Swivel Bar
Square Linkage Rods