Invacare IRC5PO2W Oxygen Equipment User Manual


 
SECTION 7—TROUBLESHOOTING GUIDE
Part No 1143482 25 Perfecto
2
Series
SECTION 7—TROUBLESHOOTING
GUIDE
SYMPTOM: PROBABLE CAUSE: SOLUTION:
Alarm:
Short beeps, long
pause
Concentrator not
operating,
power switch On.
Beep......Beep......
Main Power Loss:
1. Power cord not
plugged in.
1. Insert plug into outlet.
2. No power at outlet. 2. Inspect house circuit
breakers or fuses. If problem
recurs, use a different outlet.
3. Tripped circuit
breaker.
3. Call Service Provider.
Alarm:
Continuous
Concentrator not
operating, Power
Switch On.
Beeeeeeeep....
System Failure:
1. Unit overheating due
to blocked air intake.
1a. Remove and clean cabinet
filters.
1b. Move oxygen
concentrator at least three
inches away from walls,
draperies or furniture.
2. Insufficient power at
outlet.
2. DO NOT use extension
cords. Move to another
electrical outlet or circuit.
3. Internal repairs
required.
3. Call Service Provider.
Alarm:
Rapid
Beep....Beep...
Beep...Beep
Low Flow Alarm:
1. Kinked or blocked
tubing, cannula or
humidifier.
1. Inspect for kinks or
blockages. Correct, clean or
replace item. Once
corrected, turn power Off for
60 seconds and then turn
power back ON.
2. Flowmeter set at 0.5
L/min or less.
2. Reset flowmeter to
prescribed flowrate. If
prescribed flowrate is below
0.5 L/min, use the Pediatric
Flowmeter IRCPF16 (Only
with Perfecto
2
5).