Invacare IRC5PO2W Oxygen Equipment User Manual


 
SECTION 4—TYPICAL PRODUCT PARAMETERS
Part No 1143482 13 Perfecto
2
Series
SECTION 4—TYPICAL PRODUCT
PARAMETERS
Electrical Requirements: 120 VAC ± 10% (132 VAC/108 VAC), 60 Hz
Rated Current Input: 3 A
Sound Level: 44 ± 2 dBA Average
Altitude: Up to 6,000 ft (1828 meters) above sea level
without degradation of concentration levels.
From 6,000 ft (1828 meters) to 13,129 ft
(4000 meters) below 90% efficiency.
*Oxygen Output
Concentration Levels:
87% to 95.6% for flows ranging at 0.5 to 5 L/min
Maximum Outlet Pressure: 5 psi ± 0.5 psi (34.5 kPa ± 3.45 kPa)
Flow Range: 0.5 to 5 L/min
For flowrates less than 1 L/min, we recom-
mend the use of the Invacare Pediatric Flow-
meter IRCPF16 Accessory.
Low Flow Alarm: 0 L/min to 0.5 L/min
Rapid Audible Alarm Beeping (No
Accessories Connected).
Average Power
Consumption:
280 W (Operating @ 3 L/min)
Pressure Relief Mechanism
Operational at:
35 psi ± 3.5 psi (241 kPa ± 24.1 kPa)
Change in maximum recom-
mended flow when back pres-
sure of 7kPa is applied:
0.7 L/min
Filters: Cabinet, Outlet HEPA and Compressor Inlet
Safety System: Current overload or line surge shutdown.
High temperature compressor shutdown.
High Pressure Alarm w/compressor shut-
down. Low Pressure Alarm w/compressor
shutdown. Battery Free Power Loss Alarm.
SensO
2
Oxygen System (SensO
2
Model) Low
Flow Alarm.