Invacare IRC5PO2W Oxygen Equipment User Manual


 
SECTION 6—MAINTENANCE
Perfecto
2
Series 24 Part No 1143482
PREVENTIVE MAINTENANCE RECORD
ModelNo.IRC_______________________________
SerialNo.____________________________________
OnEachInspection
*NOTE:RefertoPreventiveMaintenancesectionofService
Manual.
Record Date Of Service
Record Elapsed Hours on
Hour Meter
Clean Cabinet Filters
Check Prescribed L/min
Flow Rate
Every 8,760 hours, During Preventive Maintenance Schedule, Or
Between Patients
Clean/Replace Cabinet
Filters
Check outlet HEPA Filter*
Check Compressor Inlet
Filter*
Check Oxygen
Concentration (SensO
2
models)
Check Power Loss Alarm
As Required
Clean Heat Exchanger
Replace Exhaust Muffler
Rebuild Top End
Compressor
Every 4,320 Hours for IRC5P
Check oxygen
Concentration