Invacare IRC5PO2W Oxygen Equipment User Manual


 
SECTION 5—OPERATING INSTRUCTIONS
Part No 1143482 19 Perfecto
2
Series
CAUTION
Neversettheflowgreaterthan5L/min.
ƽ WARNING
DONOTchangetheL/minsettingontheflowmeterunlessa
changehasbeenprescribedbyyourphysicianortherapist.
NOTE:Toproperlyreadtheflowmeter,locatetheprescribedflowrate
lineontheflowmeter.Next,turntheflowknobuntiltheballrisesto
theline.Now,centertheballontheL/minlineprescribed.
2. Iftheflowrateontheflowmetereverfallsbelow0.5L/min
forPerfecto
2
5formorethanaboutoneminute,theLOW
FLOWalarmwillbetriggered.Thisisarapidbeepingof
theaudiblealarm.Checkyourtubingoraccessoriesfor
blockedorkinkedtubingoradefectivehumidifierbottle.
Afterratedflowisrestored,morethan0.5L/minfor
Perfecto
2
5,theLOWFLOWaudiblealarmwillgooff.
NOTE:Theuseofsomeaccessoriessuchasthepediatricflowstand
andtheHomeFillcompressorwilldeactivatetheLowFlowAlarm.
FIGURE 5.5 Flowrate
Flow Knob
Ball
Flowmeter
Front Panel
0.5 L/min