ACCESSORIES
A
C
C
E
S
S
O
R
I
E
S
ACCESSORIES
Operating Instructions
1. Set the lbs./kg switch to the desired mode.
2 .With the swivel bar, chains/straps and sling
attached to the weight module, press the
OPERATE button on the front panel and adjust
to ZERO by rotating the adjustment knob until
you obtain a displayed reading of 00.0.
NOTE: The swivel bar, chains/straps and sling
can also be weighed separately, then subtracted
from the total weight which includes the patient
and attaching hardware.
3. Place the patient in the sling.
4. Lift the patient to a height sufficient enough to
ensure that the patient is not in contact with the
bed, chair, etc.
5. When the patient's motion has stabilized, press
the OPERATE button and the patient's weight
will be displayed.
NOTE: Please observe that the weight module
will automatically turn OFF after approximately
two (2) minutes to conserve power. If the
display turns OFF before you are able to record
the patients weight, simply press the OPERATE
button once again.
6. Carefully lower the patient after reading.
* HOLD Function: With the HOLD switch in the
HOLD ON position, the display reading will FREEZE
after approximately one (1) minute (lbs./kg will flash)
then automatically turn off at the end of the two (2)
minute timing cycle.
Battery Replacement
The weight module is powered by a nine (9) volt
battery. This should provide approximately 3000
readings before needing replacement.
When battery replacement is needed a B will
appear on the display.
Simply remove the two (2) screws on the sides of
the module and slide the housing upward to access
the battery.
CAUTION
Calibration and maintenance should be
performed by trained factory personnel ONLY.
The load cell transducer contains no user
serviceable components. In the event of
unauthorized tampering, warranties shall
become null and void.