Invacare 9000 SL Mobility Aid User Manual


 
9000 Series 52 Part No. 1056953
SECTION 9 RECLINER
RECLINER
INSTALLING THE TELESCOPING I.V. ROD
INSTALLING THE TELESCOPING I.V. ROD HOLDER (FIGURE 12)
1. Remove the rear wheel from the
wheelchair. Refer to
REMOVING/
INSTALLING THE REAR WHEELS in
this section of the manual.
2. Remove the locknut and hex screw
that secure the arm socket to the
wheelchair frame.
3. Remove the locknut that secures the
back tube to the clevis bracket.
4. Position the top hole on the flat
portion of the I.V. holder on the
exisitng pivot screw of the clevis
bracket as shown in FIGURE 13.
5. Reinstall locknut onto pivot screw.
DO NOT tighten yet.
6. Thread the provided mounting screw
through the lower hole on the front
cylinder section of the I.V. holder,
through the side frame, and into the
locknut. Turn nut until snug fit is
obtained. DO NOT tighten yet.
7. Reinstall the arm to orient front
cylinder of I.V. holder. Refer to RE-
MOVING/REPLACING ARMRESTS in
SECTION 4 of this manual.
8. Once front cylinder is oriented for
proper clearance of the armrest,
securely tighten both locknuts.
9. Install the I. V. rod support tube. Refer to INSTALLING THE I.V. ROD SUPPORT
TUBE in this section of the manual.
FIGURE 12 - INSTALLING THE
TELESCOPING I.V. ROD HOLDER
Hex Screw
Arm Socket (OUT-
SIDE of Wheelchair
Frame)
Locknut
STEP 2
Outside of
Wheelchair
Frame
Clevis
Bracket
(STEP 3)
Pivot Screw
(STEP 4)
Back Tube
(STEP 3)
Locknut (STEP 6)
Mounting
Screw
(STEP 6)
I.V. Holder (STEP 4)
Locknut
(STEPS 3 and 5 )