Hanna Instruments HI 96770C Thermometer User Manual


 
17
Note: It is possible to interrupt calibration procedure at any time by pressing CAL CHECK
or ON/OFF.
Warning: Do not calibrate the instrument with standard solutions other than the HANNA
CAL CHECK™ Standards, otherwise erroneous results will be obtained. For accurate Validation and
Calibration results, please perform test at room temperature: 18 to 25°C (64.5 to 77.0°F)
Turn the meter on by pressing ON/OFF.
• When the beeper sounds and the LCD
displays dashes, the meter is ready.
Enter the calibration mode by holding CAL
CHECK for three seconds. The display will
show “CAL” during calibration.
Place the CAL CHECK™ Standard
Cuvet A into the cuvet holder. Ensure that
the notch on the cap is positioned securely
into the groove.
Press ZERO/CFM and the instrument status
indicator will appear, depending on the
measurement phase.
CALIBRATION PROCEDURE