Hanna Instruments HI 96770C Thermometer User Manual


 
11
The instructions listed below should be carefully followed during testing to ensure best accuracy.
For a correct filling of the cuvet: the liquid in the cuvet forms
a concavity on the top; the bottom of this concavity must be
at the same level of the 10 mL mark.
Proper use of the powder reagent packet:
(a) use scissors to open the powder packet;
(b) push the edges of the packet to form a spout;
(c) pour out the content of the packet.
It is important that the sample does not contain any debris. This would corrupt the reading.
Each time the cuvet is used, the cap must be tightened to the same degree.
Whenever the cuvet is placed into the measurement cell, it must
be dry outside, and completely free of fingerprints, oil or dirt.
Wipe it thoroughly with HI 731318 or a lint-free cloth prior to
insertion.
Shaking the cuvet can generate bubbles in the sample, causing higher
readings. To obtain accurate measurements, remove such bubbles by
swirling or by gently tapping the cuvet.
It is possible to take multiple readings in a row, but it is recommended
to take a new zero reading for each sample and to use the same cuvet for zeroing and measurement.
All the reaction times reported in this manual are referred to 20°C (68°F). As a general rule of
thumb, they should be doubled at 10°C (50°F) and halved at 30°C (86°F).
In order to maximize accuracy, prior to a measurement follow the validation procedure to be
sure that the instrument is properly calibrated. If necessary, calibrate the instrument.
GENERAL TIPS FOR AN ACCURATE MEASUREMENT