Hanna Instruments HI 96770C Thermometer User Manual


 
13
Press ZERO/CFM and the instrument
status indicator will appear, depending on
the measurement phase.
After a few seconds, the display will show
“-0.0-”. The meter is now zeroed and ready
for measurement.
Remove the cuvet.
Add the content of one packet of
HI 96770A-0 Silica HR reagent A and
shake until complete dissolution.
Remove the cap and add the content of one
packet of HI 96770B-0 Silica HR reagent
B. Replace the cap and shake
vigorouslyvigorously
vigorouslyvigorously
vigorously
until complete dissolutionuntil complete dissolution
until complete dissolutionuntil complete dissolution
until complete dissolution.
Press and hold
READ READ
READ READ
READ
/TIMER/TIMER
/TIMER/TIMER
/TIMER for three
seconds. The display will show the
countdown. The beeper indicates the end of
countdown period.
Alternatively, wait for 10 minutes.