Golden Technologies GP215 Wheelchair User Manual


 
Alante DX/LT Owner’s Manual Rev A_011811
VII. OPERATION – cont.
1.6 Slow or Sluggish Movement
If the wheelchair does not travel at full speed or does not respond quickly enough, and
the battery condition is good, check the maximum speed setting. If adjusting the speed
setting does not remedy the problem, then there may be a non-hazardous fault. Contact
your Golden Technologies representative.
1.7 Maximum Speed/Profile Indicator is Steady
The display will vary slightly depending on whether the control system is programmed to
operate with drive profiles.
1.7.1 Maximum Speed Indication
The number of LEDs illuminated shows the maximum speed setting. For example, if the
setting is speed level 4, then the four left hand LEDs will be illuminated.
1.7.2 Profile Indication
The LED illuminated shows the selected drive profile. For example, if drive profile 4 is
selected, then the fourth LED from the left will be illuminated.
1.8 Maximum Speed/Profile Indicator Ripples Up and Down
This indicated the control system is locked.
1.9 Maximum Speed/Profile Indicator Flashes
This indicates the speed of the wheelchair is being limited for safety reasons. The exact
reason will depend on the type of wheelchair; however, the most common cause is that
the seat is in the elevated position.
The VR-2 modules are not user serviceable. Do not attempt to repair VR-2
modules.
Note that joystick OONAPU (Out Of Neutral At Power Up) is not a fault. Simply by
removing your hand from the joystick and allowing it to return to the neutral position,
the fault will immediately clear. If the condition persists after removing your hand, the
joystick may be damaged. Consult a Golden Technologies representative.
30