3. TO REMOVE A THAWED BAG
3.1 Press the ROCK key to stop the rocking motion.
a. Remove the thawed bag (and any weights used) from the
Plasma Bag Holder. Discard plasma thawing overwrap bag
after each use, as they are designed for single use only.
b. If there is another bag being thawed, press the ROCK key to
resume the rocking motion for the remaining bag.
4. TO MAINTAIN THE DEVICE IN THE READY-TO-
THAW STATUS
4.1 Always keep the HEAT indicator illuminated.
4.2 Always keep the SET TEMP to the desired setting. (37 °C or
other desired temperature). If a desired temperature of 50
°C or greater is needed, limit the time the unit is in the ready
to use state. At higher temperatures, greater evaporation
occurs.
4.3 Always keep the ROCK indicator off (not illuminated) when
not thawing plasma or other products.
9