Genesis Advanced Technologies 2002 Bathroom Aids User Manual


 
2) The ADD WATER light is on and there is sufficient
water in the bath.
With the unit unplugged and raised out of the water, check
that there is no obstruction around the float switch in the
shroud unit. The float should move freely up and down the
stem approximately half-inch.
3) The FAULT light is on.
Check the water temperature. If the water temperature is
above 58 °C, the FAULT light will light and the unit will
not be operable.
If this temperature is reached, it may indicate a unit
malfunction.
Contact
GenesisBPS, LLC for further assistance.
4) Unit does not heat. HEATING light is on but actual
water temperature is not increasing.
Remove rear cover and check the heating element for
continuity. Replace if necessary. This maintenance should
only be performed by a trained technician or
GenesisBPS,
LLC. Contact it or your Biomedical Department.
5) Unit does not rock
Check that the ADD WATER or FAULT lights are not
illuminated.
6) Bath moves to side when rocking or the rocking motion is
not smooth.
Ensure that the bath is positioned on the white roller.
The unit must be operated on a smooth and level surface.
Check the level of the work surface and make adjustments
if necessary.
16