Genesis Advanced Technologies 2002 Bathroom Aids User Manual


 
indicator will light. To correct this situation, fill bath with
water as explained above.
Add 2 ml of LABath CLEAN
TM
each time the water is
changed to control bacterial and algae growth. The bath
cleaning agent also helps to prevent the formation of turbidity
and odor.
2.3 Plug the device into a 115 volt AC outlet. For GENESIS
2002, a dedicated 10 Ampere outlet is needed. For the
GENESIS 2006, a dedicated 15 Ampere outlet is needed.
The outlet must be properly grounded.
IMPORTANT : DO NOT use sharp instruments (pen,
pencil, etc.) on the key pad area - it will cause damage
to the instrument.
2.4 Turn on the POWER SWITCH on the left panel.
2.5 Start the unit by pressing the POWER key located on the
control panel.
a. The default SET TEMP is 37 °C. When switching on the unit
this temperature will be displayed in the SET TEMP window,
as the desired temperature, along with the actual WATER
TEMP of the bath. The bag timers will display zeros.
b. If the water temperature is less than the set temperature
(37 °C) then heating of the water will begin automatically -
the HEAT indicator will light. In addition, the HEATING
indicator will light, indicating the heater is drawing power.
When the bath water reaches the set temperature, HEATING
indicator will go off. When the bath water cools, HEATING
indicator will go on again, indicating water is being heated.
This cycle continues, ensuring accurate water temperature.
6