The over-temperature cut-out should be checked every 6 months to see
that it is working properly. In order to check the user selected cut-out, fol
-
low the controller directions (Section 8.2) for setting the cut-out. Both the
manual and the auto reset option of the cut-out should be checked. Set the
instrument temperature higher than the cut-out. Check to see if the display
flashes cut-out and the temperature is decreasing.