Emerson Process Management 430 DO Oxygen Equipment User Manual


 
2
MODEL 430 DO SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
1.3 AVAILABLE MODELS
The Model 430-01 and -02 Sensor is a flow through
assembly. The sensor is supplied in a 1-1/2 inch PVC
tee for installation into a process line.
The Model 430-03 Submersible Sensor is for use in
those applications where the required flow velocity can
be maintained without the use of an agitator. When the
measurement is being made in a stagnant pond or
other media where the minimum flow rate of 1.5 FPS
cannot be obtained, the Model 430-04 must be select-
ed. This model is designed for submersible use and is
provided with its own agitator to ensure adequate flow
velocity at the membrane.
The Models 430-03 and 04 may be provided with an
optional mounting bracket assembly. This bracket is
constructed of light weight aluminum and can be
adjusted to satisfy virtually any aeration basin mount-
ing requirement. Once installed, the probe can be
removed without the use of tools. A single locking pin
holds the assembly rigidly in place but allows it to be
easily removed for inspection or cleaning.
1.4 SPECIFICATIONS
Process Connection:
430-00/01 1-1/2 inch socket fitting
Wetted Materials: HOUSING - PVC
MEMBRANE - Teflon
PRESSURE COMPENSATOR - Neoprene
Temperature Compensation:
0 to 50°C (32 to 122°F) Automatic
Sample Pressure: 0-50 psig (345 kPa abs)
Submersible to 200 ft. (61 m).
Agitator Power Requirement: Model 430-04 only)
115 Vac, ±10 60 Hz (0.2 amp)
Cable Length (standard):
6.1 meters (20 ft) on sensor and agitator
Maximum Cable Length:
305 meters (1000 ft)
Operating Time Between Electrolyte Recharge:
Approximately 12 months or longer, depending upon
application
Sample Flow Requirements for Models 430-01-02-03:
1.5 feet per seconds (0.5 m/s)
FIGURE 1-1 Flow Through Assemblies
DWG. NO. REV.
40043019 A
MILLIMETER
INCH