Citizen CT503 Scale User Manual


 
8
Error Codes
Error codes are shown in the main display for approx. 2 seconds, after which the program automatically returns to the
weighing mode.
Display Cause Solution
No segments appear on No AC power is available Check the AC power supply
the display The AC adapter is not Plug in the AC adapter
plugged in
Battery is dead Replace the battery
Recharge the battery using
an external recharging device
h The load exceeds the Unload the scale
scale capacity
l The load plate is not in place Place the load plate on the
scale
Something is touching Move the object that is
the load plate touching the load place
e 01 Display capacity exceeded: Decrease the weight on
Value to be output cannot be the scale
shown on the display
e 02 Calibration parameter Calibrate only when zero
not met; e.g.: is displayed
– scale not zeroed Press t to tare the
scale
– scale is loaded Unload the scale
e 09 When gross - zero, no tare Tare the scale
e 10 The t key is blocked Press [CAL/CF] > 2 sec. to clear
when there is data in the tare memory and release the
the second tare memory tare key
(net-total) – only 1 tare function
can be used at a time
e 11 Value input is not allowed for Press t
second tare memory
e 22 Weight is too light or there is Increase the reference quantity
no sample on the balance
e 30 Interface port for printer Contact your local
output is blocked Service Center
Max. weighing range is less The scale was Place the weighing pan on
than indicated under turned on without the weighing on the scale
”Specifications” pan in place and press [ON/OFF] to turn on the
scale
The weight readout The scale has not Calibrate/adjust the scale
is obviously wrong been calibrated/adjusted
The scale was not Tare before weighing
tared before weighing
If any other errors occur, contact your local Service Center!