When the measurement has been completed, the systolic
blood pressure, the diastolic blood pressure and the pulse
rate are alternately displayed and the air is discharged.
The systolic blood pressure and the diastolic blood pres-
sure are displayed for about 2 seconds and the pulse is
displayed for about 1 second.
The measured data are registered automatically, as long as
they are not error.
When a person with maximum blood pressure of 170 mmHg
or higher uses the unit for the first time, or after changing
batteries, measure blood pressure using the manual set-
ting. If the automatic pressure setting is used, it may not
work.
Automatic Pressure Setting
This product automatically sets the pressure based on
the previous measurement, so you can easily measure
blood pressure by pressing one button without reset-
ting the pressure for each measurement. If several people
use the unit, we recommend using the manual settings
as follows.
Manual Pressure Setting
When several people use the unit, set the pressure value
manually. Please note that if the manual pressure set-
ting is not used, pressure will be automatically applied
based on the value set by the previous user.
If the previous user had a high pressure setting, that
high pressure will automatically be used for the current
user. If you feel pain in your arm, press the POWER /
START switch immediately to turn of the power.
1. Put the arm cuff around your arm, plug the air hose
into the unit, and keep your arm at the level of your
heart.
2. Keep the POWER / START switch depressed until
the pressure display shows the optimal pressure set-
ting. (The optimal pressure setting is about 40 mmHg
higher than your usual maximum blood pressure.)
3. When the pressure display reaches the optimal pres-
sure value, release the POWER / START switch. When
the POWER / START switch is released pressuriza-
tion will finish and measurement will start. Manual pres-
sure settings can be selected between 160 mmHg - 280
mmHg. If you release the POWER / START switch
before reaching 160 mmHg, the unit will pressurize to
160 mmHg. If the unit senses that the pressure is not
high enough, it will apply more pressure.