Bayer HealthCare Ascensia BREEZE and Ascensia AUTODISCTM Unique 10-Test Disc Blood Glucose Meter User Manual


 
Comparison of Meter to Laboratory Results
† Weighted Deming regression, weighted linear correlation coefficient
Opr. = Operator; Lay = Lay User; HCP = Healthcare Professional
36
0.0
5.0
10.0
15.0
20.0
25.0
30.0
35.0
0.0 5.0 10.0 15.0 20.0 25.0 30.0 35.0
Laboratory Plasma Glucose Result (
mmol/L)
Ascensia
®
BREEZE™ Result (mmol/L)
o = Lot A + = Lot B
y=x
95% Confidence Intervals
Lot Opr. N Y= Slope / Intercept R
A Lay 100 0.85x + 0.72 0.81
to
0.89 0.38
to
1.06 0.98
HCP 100 0.84x + 0.65 0.80
to
0.88 0.33
to
0.96 0.98
B Lay 96 1.02x – 0.58 0.96
to
1.08 –1.04
to
–0.13 0.97
HCP 96 1.03x – 0.61 0.99
to
1.08 –0.94
to
–0.29 0.98
A&B All 392 0.91x + 0.21 0.89
to
0.94 0.01
to
0.40 0.97
B. Healthcare Professional Results