CALIBRATE MILLIAMP INPUTS
mA, mA % (Percent of 4 to 20 mA), DP% (DP Flow)
Choose this function to provide an output from 0.00 to 24.00
milliamps. The compliance voltage is a nominal 24 VDC to provide
the driving power to your milliamp receivers.
1) Disconnect one or both input wires from the device to be
calibrated.
2) Press the MILLIAMP/V push-button until mA or mA and %
appear on the display
3) Press the DISPLAY/SOURCE/READ push-button until SOURCE
or 2-WIRE appear on the display
4) If 2-WIRE is on the display, press the TYPE/ENG UNITS push-
button once to indicate SOURCE on the display
5) Press the mA/%/% DP FLOW push-button to display mA, % 4-20
or % DP Flow.
6) Connect the red SOURCE lead of the calibrator to the plus (+)
input of the device and the black SOURCE lead to the minus (-).
Output current is continuously adjustable with the UP/DOWN
push-buttons. Zero & Span (or any other values) are available by
using the LO and HI "QUIK-CHEKs".
OPERATING INSTRUCTIONS
Milliamp Receiver Input
Controller
Transmitter
Computer
Logger
I/P
DCS
SOURCE MODE
Select source by pressing the DISPLAY/SOURCE/READ
pushbutton until the word SOURCE appears on the LCD display.
To change the output value, press the UP or DOWN key for each
digit. To ramp the output press and hold any UP or DOWN
pushbutton. The display will continue to change in increments
corresponding to the digit being changed and will automatically
carry up and down until the limits of the range are reached. This
function operates in all three output positions (HI, SET & LO).
STORING QUIK-CHEK OUTPUTS
1) Press HI or LO
2) Press the UP/DOWN keys to desired
value
3) Press the STORE push-button
The LCD will ash once to show that the
value was saved
If a value is in the SET position and you want
that value stored in HI or LO, press and hold
the STORE push-button, then press the HI
or LO push-button. The display will ash
once to indicate the value has been stored.
Then release both push-buttons.
RECALLING QUIK-CHEK OUTPUTS
When you need a stored value just press a
Quik-Chek push-button. Any value for the
selected range may be stored in HI & LO.
The TechChek 830-KP remembers the HI,
LO and SET values for each function with the
power on or off. Each time a different function
is selected, the last three QUIK-CHEK values
for that function will be recalled.
OPERATING INSTRUCTIONS
8