Agilent Technologies 1141A Thermometer User Manual


 
40
Calibration Tests and Adjustment
Probe Adjustment
Adjustments
This section provides adjustment procedures for the 1141A Differential Probe and
attenuator adapters. There are no service adjustments for the 1142A Probe Control
Module.
Adjustment Interval
None of the adjustment procedures that follow should be considered for a routine
maintenance plan. The differential probe and attenuator adapters should be adjusted
under conditions specified at the beginning of the respective procedures.
NOTE Warm up the instrument for 30 minutes before starting adjustment procedures.
Probe Adjustment
This procedure adjust the high-frequency and low-frequency paths on the 1141A
Differential Probe.
NOTE Do not perform this procedure as a part of routine maintenance. Perform the procedure
only if the probe does not meet specifications or has been repaired.
CAUTION You are going to remove the covers of the probe, so the assembly inside will be exposed
while under power. The PC assembly will be electrically and mechanically vulnerable.
Do these adjustment procedures in an ESD-safe area.
Avoid inadvertent contact between the powered assembly and nearby tools and
equipment.
Avoid mechanical damage by carefully handling the exposed assembly and cables.