CHAPTER 4
Using terminal commands
43
— 4 for A733
— 5 for A723
— 6 for A440
• slot and samples are the actual values programmed by
means of the SLOT command
• po (A723 only) is the relative output power
• err_level is the error value; 0 means no error
REMARKS GET only.
REMOTE Yes, GET only. The A720B and A723 can issue the command both
remotely and locally, while the A720 can issue the command only
locally.
EXAMPLE INFO
193 INFO 255 0 18/4/1999 21:5:11 1.3 0 0 0 91 72
40 1:46 58 65 72 3 900 0 0
#
COMMAND ANLG
APPLIES TO A723
DESCRIPTION Sets/returns various parameters of the analog subsystem (e. g. the
sampling/averaging method used for individual data aquisition
channels).
PARAMETERS A control byte specifying the command and the analog input chan-
nel number the command is acting on:
Figure 13. The ANLG Control Byte Layout.
Some commands may also require one or two additional 16 bit
parameters representing threshold values.
• The Channel Number selects the analog channel that will be
affected by the command. For the A723 device, only 0000 to
0101 are accepted (only 6 analog channels are available).
Command Code
Channel Number